Service Manual Indicator JIK-6 Page 1
Chapter I Basic Introduction
1-1 Dimension------------------------------------------------------------------------------- 1
1-2 Cross-reference List of Characters--------------------------------------------------- 1
1-3 Product Specification Sheet----------------------------------------------------------- 2
Chapter II Parameter Function Setting
2-1 General Parameter Setting------------------------------------------------------------ 3
2-2
Advanced Parameter Setting--------------------------------------------------------
--
5
2-3 Weighing Capacity, Sensed Weight Setting and Weight Calibration------------ 7
Chapter III Various Assemblies and Relevant Illustrations
3-1 Back Cover Assembly and Disassembly-------------------------------------------- 9
3-2 Installation of Additional Board------------------------------------------------------ 10
3-3
Wiring Cross-reference Chart--------------------------------------------------------- 11
3-4
Illustration of Contacts of Additional Board---------------------------------------- 12
Chapter IV Troubleshooting
4-1
Preliminary Examination-------------------------------------------------------------- 13
4-2
Error Message and Troubleshooting----------------------------------------------
---
13
4-3
Summary Table of Error Message-------------------------------------------------
---
16
Chapter V
Structural Part of Platform Scale
5-1
Connection of Metal Connectors---------------------------------------------------
---
17
5-2
L/C Replacement Process------------------------------------------------------------
-
17
5-3
L/C-related Technology--------------------------------------------------------------
--
19
Chapter VI Exploded Drawing and Cross-reference List of Parts
6-1
JIK 6-CAB ------------------------------------------------------------------------------ 20
6-2
JIK 6-CSB ------------------------------------------------------------------------------ 21
6-3
JIK 6-CSN ------------------------------------------------------------------------------ 22
Vetek AB
Hantverksvägen 15, 76493 Väddö - Sweden
Tel_ +46176-208920, info@vetek.com, www.vetek.com
Ser
vice_Manual_JIK-6_V1

Questions and Answers

  • J
    Jaime WellsSep 10, 2025
    What does error E10 mean on Vetek JIK-6 RS232 (RTC) Low Battery?
    • S
      Stephen SotoSep 10, 2025
      The error E10 on Vetek Measuring Instruments indicates RS232 (RTC) Low Battery, it can be caused by a battery problem or a board problem. Check if the RTC battery is low or if there is bad contact. Also, check if the RS232 board and main board have a bad connection. If the problem continues, replace the board.
  • C
    Carrie HullAug 22, 2025
    What to do if my Vetek JIK-6 Measuring Instruments has linear defect or incorrect weighing?
    • M
      Miranda AdkinsAug 22, 2025
      If your Vetek Measuring Instruments is showing a linear defect or incorrect weighing, this could be due to a wrong internal setting, a three-point calibration error, interference from an unknown object, a defect in the L/C, or a defect in the A/D board. To solve the problem: * Refer to “Various Parameters Setting” and re-enter the correct setting values. * Set the maximum and minimum weighing capacity according to “Weighing Capacity, Sensed Weight and Weight Calibration” and perform a three-segment calibration after setting is completed. * Inspect the exterior and interior of the scale to see if something is interfering with the L/C structure and the weighing pan. * Replace the L/C directly to see if it is defective. * Replace the A/D board directly.
  • B
    Bradley WarnerAug 25, 2025
    How to fix Vetek JIK-6 calibration failure?
    • J
      Jennifer FloresAug 25, 2025
      A calibration failure in Vetek Measuring Instruments can stem from a wrong internal setting, interference from an unknown object, a defective L/C, or a defect in the A/D board or display panel. Try resetting the device using the “Various Parameters Setting”. Inspect the scale's exterior and interior for any obstructions to the L/C structure or weighing pan. If the problem persists, consider replacing the L/C, A/D board, or the display panel directly.
  • P
    Peter RiosAug 27, 2025
    What causes unstable zero point on Vetek Measuring Instruments?
    • D
      David MooreAug 27, 2025
      An unstable zero point on your Vetek Measuring Instruments can be caused by interference from the external environment, a three-point calibration value error, interference from an unknown object, or a defective L/C. To address this, first, eliminate any potential environmental interference. If the issue persists, recalibrate the device. Also, inspect the exterior and interior of the scale for any obstructions to the L/C structure or weighing pan. If these steps don't resolve the problem, consider replacing the L/C.
  • A
    Amanda CarrilloAug 27, 2025
    What to do if Vetek JIK-6 shows 'E2 Initial Zero Point Too High' or 'E3 Initial Zero Point Too Low'?
    • H
      hansonvickiAug 27, 2025
      If your Vetek Measuring Instruments displays 'E2 Initial Zero Point Too High' or 'E3 Initial Zero Point Too Low', it could be due to several reasons: the protection screw of the structure not being removed, interference of the overload protection screw, an object not completely removed from the weighing pan during power on, interference from an unknown object, or the offset value differing from the factory setting due to transportation or other factors. Check and make appropriate adjustments. Inspect the area around the weighing pan. Perform a three-segment calibration to re-record the new initial zero point value.
  • S
    Sarah GarciaAug 29, 2025
    How to troubleshoot poor LCD display on Vetek JIK-6?
    • C
      Christina HickmanAug 29, 2025
      A poor LCD display, characterized by broken or missing strokes on your Vetek Measuring Instruments, may be due to an LCD defect, a CPU welding joint defect, or a display panel defect. First, ensure there are no broken board lines or short circuits in the connection between the LCD and other conducting objects. If not, replace the LCD directly. Also, re-weld the contact of the program IC (pos. no. U8) on the main board using a soldering iron. If the issue persists, replace the display panel directly.
  • D
    Diana FrenchSep 18, 2025
    What does error E21 mean on Vetek Measuring Instruments?
    • T
      thompsonsarahSep 18, 2025
      The E21 error on Vetek Measuring Instruments appears when the basic model’s e resolution ratio is less than 100, or when the advanced model’s e resolution ratio is less than 150, based on OIML standards. Reset the settings according to “Weighing Capacity, Sensed Weight and Weight Calibration”.
  • T
    tiffany25Sep 12, 2025
    What does E11 error mean on Vetek Measuring Instruments?
    • D
      davisjanetSep 12, 2025
      The E11 error on Vetek Measuring Instruments means that in accumulation mode, data of different units can’t be saved (e.g., Kg and pcs); only one fixed unit can be used at one time. Clear all accumulation data by pressing the key twice, then press the key to return to normal weighing mode.
  • T
    Tyler KerrSep 8, 2025
    What does E6 error mean on Vetek JIK-6 Measuring Instruments?
    • G
      garrett97Sep 8, 2025
      The E6 error on your Vetek Measuring Instruments indicates bad contact of metal connector, L/C line damaged or broken, or L/C problem. Clean male and female sockets of the metal connector. Check if any external force, insect or rat damages L/C line. You may also need to address L/C-related issues or consider replacing the L/C.
  • D
    Darlene BaileySep 5, 2025
    How do I fix the 'E1 No calibration data' error on Vetek JIK-6?
    • A
      Angela DominguezSep 5, 2025
      The 'E1 No calibration data' error on your Vetek Measuring Instruments indicates a calibration setting error. To fix this, recalibrate the device.