6.6. INITIAL START-UP
6.6.1. Filling the Water Tank
To ll the water tank, remove the cap underneath the slider of the water tank. Carefully ll the tank with
distilled or deionized water.
The conductivity of the water used in the generator must not exceed 2μS/cm.
Fill the tank to the maximum level indicator. Close the slider cap.
Do not ll the water tank higher than up to the maximum mark.
To prevent contamination of the cell assembly, it is important to use only deionized
or distilled water in the generator. Water containing metallic impurities will
contaminate or damage the cell and will void the warranty.