9. MAINTENANCE
With proper care and maintenance, your hydrogen generator should provide you with years of trouble-free
operation. There are no adjustments to be made to the generator. The only routine service operations are
those described below.
Nonetheless, the generator should be inspected approximately every 2 years. Contact your supplier.
9.1. ROUTINE MAINTENANCE
The following section describes the maintenance operations required for the correct operation of the
hydrogen generator.
9.1.1. Clea ning
The internal components of the hydrogen generator do not need to be cleaned and should not be
accessed by the user for cleaning. To clean the outside of the unit, only use a damp cloth (no detergents,
acids or aggressive or abrasive substances).
9.1.2. Water Relling
9.1.3. Deionizer Bag Replacement
The tank must be relled when the water level approaches the lower level, and the “Rell Water” pre-
alarm message appears. Use distilled or deionized water.
Rinse the water tank and replace the 2 deionizer bags approximately every six months, or whenever the
“Change Water” message appears.
The part-number for the deionizer bags is: DB-H200-031 (includes 2 bags). See section 6.2.2 for more
details.