19617-001 Rev. H DiagnosticPRO Edge User’s Guide 27
Appendix:
Electromagnetic guidance
Caution: Medical electrical equipment.
EMC (Electro Magnetic Compatibility) must be considered before any medical electrical equipment is installed or
put into service. Follow the information in the accompanying documentation when installing and operating the
DiagnosticPRO Edge Digitizer.
Caution: Portable or mobile RF communication equipment can effect Medical Electrical equipment.
Caution: Using the Digitizer adjacent to or stacked with other equipment may cause interference between the
equipment. Before utilizing stacked or adjacent equipment, verify proper functionality of all equipment in the
actual configuration in which it will operate.
Caution: Connecting the Digitizer to equipment that is not rated CISPR 11 class A or class B may alter the
electromagnetic characteristics.
Caution: In a domestic environment this product may cause radio interference in which case the user may be
required to take adequate measures.
Guidance and manufacturer’s declaration –
electromagnetic equipment
The DiagnosticPRO Edge Digitizer is intended for use in a professional healthcare facility environment. The customer
or end user of the Digitizer should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF Emissions
CISPR11
Group 1
The Digitizer uses RF energy only for its internal functions.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class A
The Digitizer is suitable for use in all establishments other than
domestic and those directly connected to the public low voltage
power supply network that supplies buildings used for domestic
purposes.