Vitalograph Spirotrac Instructions for Use
IFU Part No.: 07402
Issue 22
_______________________________________________________________________________________________________________
Page 53 of 78 DT_0006-15
3. Recording time: minimum of 10 seconds, maximum of 255 seconds. Default 10
seconds.
4. The number of ECG analyses that a user may select (between 1 and 8). Default
3.
5. Parameters available for selection are:
a. QTcF : Default on
b. QTcFr: Default off
c. QTcB : Default off
d. QTcH : Default off
6. The following report formats are selectable:
a. 3x4
b. 6x2
c. 12x1
d. Cabrera
e. Average beat cycle
4.6.6 Glasgow Interpretation Algorithm
The Glasgow Program has been incorporated into the Vitalograph Model 7000 and is
intended to provide an interpretation of the resting 12 lead ECG in all situations,
whether this be in a hospital or primary care setting. This algorithm has been reviewed
against the industry standard, the CSE (Common Standards for Quantitative
Electrocardiography) Database.
4.6.7 ECG Fault Finding Guide
Malfunction Cause of error Rectification
The Bluetooth driver
requests the entry of a PIN.
The device was decoupled and
is trying to establish a fixed
connection.
Enter “1111” and confirm this input.
One or several electrode
contact points are flashing
despite the electrodes being
applied.
Electrode contact is poor.
Check the positioning and the skin
contact of the electrodes.
Press the electrodes firmly on the
skin. Shave any hair or clean the
skin. Use abrader tape. Replace the
electrode. If there is no
improvement, the cable might be
defective. Please send the device
back to the manufacturer for repairs.
A beep sounds:
2 seconds on, 2 seconds off,
for 1 minute.
Wireless connection is
interrupted.
Reduce the distance to the receiver
unit or remove any existing obstacle.
Beeps at the same
frequency as the flashing of
the electrode contact points.
Electrode has probably come
loose during measurement.
Check the electrode contact.
Replace the electrode, if necessary.
“Connection error” message
The PIN was not correctly
Reconnect with the device and enter