The main display (3) on the display shows the manometer pressure or the differential pressure up to 5,000 Pa. You can choose
between the ve measuring units psi, mbar, Pa, inH
2
O and mmH
2
O for the display.
Push “P/V/F” once to switch to the pressure mode (PRESS). Push the button “Unit
” several times if required to select the desired
measuring unit.
Connect a hose to the “+” input of the pitot tube anemometer and leave the “-” input open.
The device registers the ambience conditions via the hose opening. Push and hold the button “Hold/Zero” for 2 seconds to perform
a zero reconciliation.
Place the open end of the hose in another environment than the anemometer.
The anemometer now shows the pressure difference between the “+” input (ambience conditions at the hose end) and the “-” input
(reference ambience conditions at the anemometer).
"+"-Input