36
Resetting your 9150
If the above troubleshooting tips do not correct the problem the following might help.
Restart the Meter by pressing the on/off key (2 sec).
Switch connections between a faulty input and the nearby input a see if the problem follows the load cell
or the input. Do forget to switch back the connections when the test is done.
Replace the Smart box. Use the restore function (Maintenance) to give the new Smart box the same
configuration and calibration as the old one.