EasyManua.ls Logo

Waltron 3041 - Figure 2.3: Color Development

Waltron 3041
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Introduction
17
Waltron User Manual 101-016-B.3
3041 Silica Analyzer
Absorbance for liquids is defined as the negative logarithm of the transmittance:
A = - log
10
T = log
10
1/T = log
10
I
0
/I
1
I
0
= light intensity through the sample before colorimetric reaction
I
1
= light intensity through the sample after colorimetric reaction
In most cases the absorbance has a linear correlation with the sample concentration; for a linear
calibration line just the blank and span values (i.e. concentration with zero analyte and the
maximum expected concentration) are needed. Multiple analyses at each point are averaged to
provide a more accurate calibration line.
The absorbance ranges from 0 to 1, but it can go higher.
An absorbance of 0 at some wavelength means that no light of that particular wavelength
has been absorbed. The intensities of the sample and reference beam are both the same, so
the ratio I
0
/I
1
is 1. Log
10
of 1 is zero.
An absorbance of 1 occurs when 90% of the light at that wavelength has been absorbed
which means that the intensity is 10% of the blank value. In that case, I
0
/ I
1
is 100/10 (=10)
and log
10
of 10 is 1.
Absorption photometry (Colorimetry):
The method used is based on the formation of a colored complex of the analyte with a reagent.
Light with a specific wavelength is passed through the sample and reagent solution. . The
absorbance of this light by the reaction complex, measured by a photometer, is related to the
concentration of the analyte.
readingsensor
reference
Absorbance log
Figure 2.3: Color development.

Table of Contents