EasyManua.ls Logo

Waltron 3041 - Program Menu; Settings Menu; Figure 5.7: User Interface after Pressing the PROGRAM Menu; Figure 5.8: the SETTINGS Menu and Options Are Displayed

Waltron 3041
55 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Interface
37
Waltron User Manual 101-016-B.3
3041 Silica Analyzer
5.8 PROGRAM MENU
Figure 5.7: User interface after pressing the PROGRAM Menu.
The PROGRAM Menu allows access to the following operations:
SETTINGS:
Allows the user to set the ratio of the analysis cycles versus the extra cycles, the alarm value
(low or high) and to enable or disable the loss of sample alarm and the 4-20 mA output.
CALIBRATION MENU:
This function is protected with the 1st level of password. Allows the user to set the value of
the standard used for calibration and to perform blank and calibration cycles.
5.8.1 SETTINGS MENU
Figure 5.8: The SETTINGS Menu and options are displayed.

Table of Contents