Why does Waters Laboratory Equipment reproducibility fail to meet specifications?
- Ddawn28Aug 4, 2025
Several factors can cause reproducibility issues with Waters Laboratory Equipment. These include air in the flow path (perform an initial wash using the 2707 Console or control panel), a leaking syringe (check syringe mounting and gasket; replace if leaking at the bottom), a leaking syringe valve (inspect and replace if needed), a worn rotor seal (replace the seal and inspect the stator), dead volumes in tubing connections (redo connections with new ferrules and nuts), or a broken degasser (repair and confirm operation).