Waters 2707
Autosampler
Operators Guide
71500168202 / Revision A
Copyright © Waters Corporation 2008
All rights reserved

Table of Contents

Other manuals for Waters 2707

Questions and Answers

  • C
    Charles JohnstonAug 4, 2025
    Why does Waters Laboratory Equipment reproducibility fail to meet specifications?
    • D
      dawn28Aug 4, 2025
      Several factors can cause reproducibility issues with Waters Laboratory Equipment. These include air in the flow path (perform an initial wash using the 2707 Console or control panel), a leaking syringe (check syringe mounting and gasket; replace if leaking at the bottom), a leaking syringe valve (inspect and replace if needed), a worn rotor seal (replace the seal and inspect the stator), dead volumes in tubing connections (redo connections with new ferrules and nuts), or a broken degasser (repair and confirm operation).
  • A
    April MullenAug 16, 2025
    How to resolve error messages on Waters Laboratory Equipment?
    • S
      Steven ThomasAug 16, 2025
      When Waters Laboratory Equipment displays an error message due to a fault, try system initialization or reset. You can reset the system using the Control menu of the 2707 Console.
  • J
    Jennifer AcostaAug 22, 2025
    What to do if a blank gives a peak too high for the criteria in Waters Laboratory Equipment?
    • C
      Casey ParkAug 22, 2025
      If a blank gives a peak that is too high, it might be due to a solubility problem. In this case, modify the sample or wash solution. Alternatively, the blank could be contaminated, so try using a new blank.
  • A
    Andrea MayAug 28, 2025
    Why do power surges affect Waters Laboratory Equipment?
    • J
      Jennifer PetersenAug 28, 2025
      Power surges, line spikes, and transient energy sources can disrupt the operation of Waters Laboratory Equipment. To prevent this, ensure that the instrument’s electrical supply is adequately protected from these conditions and properly grounded.
  • G
    gomezmikeNov 8, 2025
    What to do if no injection occurs with Waters Laboratory Equipment?
    • J
      Jeff ReevesNov 9, 2025
      If no injection occurs, it could be due to a blockage in the flow path. To resolve this: 1. Disconnect the needle from the valve. 2. Start a manual wash. 3. If solvent flows from the injection port, inspect the needle; if no solvent flows from the injection port, disconnect the buffer tubing from the injector valve. 4. Start a manual wash. 5. If solvent flows from the open end, inspect the rotor seal; if not, disconnect the buffer tubing from the syringe valve. 6. Start a manual wash. 7. If solvent flows from syringe valve, inspect the buffer tubing; if not, check for overtightened connections throughout the flow path and check the syringe valve. Another cause may be leakage in the injection valve. To resolve this: 1. Disconnect the needle tubing and buffer tubing. 2. Connect port 1 to an ...

Summary

Safety considerations

Safety advisories

Operating this instrument

1 Understanding Operating Principles

Exploring injection principles

Explains the different injection modes (full loop, partial loop) and their mechanisms.

2 Preparing the Autosampler

Plumbing the instrument

Details connecting tubing, filling the wash bottle, and accessing instrument components.

Connecting to the electricity source

Instructs on safely connecting the instrument to a grounded electricity source and powering it on.

3 Maintaining the Autosampler

Heeding maintenance considerations

Covers essential safety and handling precautions to follow during instrument maintenance.

Replacing the sample needle

Provides a step-by-step procedure for replacing the instrument's sample needle assembly.

Replacing a syringe

Details the procedure for replacing the instrument's syringe, including disconnect and reconnect steps.

4 Troubleshooting

Conducting analytical troubleshooting

Provides methods for identifying and resolving analytical problems like poor reproducibility or carryover.

A Safety Advisories

Waters 2707 Specifications

General IconGeneral
BrandWaters
Model2707
CategoryLaboratory Equipment
LanguageEnglish