2-14 Preparing the Autosampler
Installing the prep kit
Installing the optional prep kit involves removing and replacing the following
accessories:
Injector valve with sample loop
Tip: The 10 mL prep loop is large.
Syringe (2500 µL instead of 500 µL)
Buffer tube (2000 µL instead of 1000 µL)
Needle assembly (60 µL instead of 15 µL)
Air needle (for 10-mL vials)
Vial trays (for 10-mL vials)
Requirement: To install the prep kit, from the 2707 Console Configure
menu select the Prep mode check box on the Volumes tab. Then select
the Use Prep Mode check box.
Tip: For procedures on removing and replacing prep kit accessories, see
Chapter 3.
Installing the biocompatibility kit
Installing the optional biocompatibility kit involves removing and replacing
the following accessories:
Injector valve with sample loop (PEEK instead of stainless steel)
Needle assembly (silica-coated stainless steel version instead of
stainless steel)
Inlet (pump) and outlet (column) tubing with plastic fittings (not
included in the startup kit)
Caution: To avoid damaging the manifold, avoid using metal fittings in
the biocompatible injector valve.