EasyManua.ls Logo

Waters 2707 - Replacing the sample needle

Waters 2707
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-6 Maintaining the Autosampler
Recommendation: Tighten each screw a half-turn each in rotation until
all screws are tight.
7. With port 1 pointing upward, hold the valve for mounting.
8. Place the valve in its slot, and fasten it.
9. Reconnect all tubing to the valve.
10. Perform a needle wash by right-clicking in the Run Sample control panel
or via the Console Control menu.
Installing a sample loop
The instrument is fitted with a 100-µL sample loop as standard equipment.
You can install a different sample loop provided you select the proper
combination of syringe and tubing.
Connect the loop between ports 2 and 5 of the injection valve. When installing
a loop of nonstandard volume, from the Configure menu of the 2707 Console
click Volumes... and adapt settings.
Tip: Calculate maximum injection volumes using the following
equations:
Full loop – Injection volume = loop volume
Partial loop needle overfill – Maximum injection volume = 50% of
loop volume
Partial loop – Maximum injection volume = (loop volume - [3 ×
needle volume])/2
Replacing the sample needle
To replace the sample needle
1. From the Maintain menu in the 2707 Console, click Replace and the
needle.
Result: The needle moves to the exchange position.
2. Follow the instructions in the Replace Needle warning box, and then
click OK.
3. Loosen the needle connection nut.

Table of Contents

Other manuals for Waters 2707

Related product manuals