C-2 Specifications
Outlining general specifications
General
Sound pressure level < 65 dB
Working temperature 10 - 40°C (indoor use only)
Storage temperature –25 - +60°C
Humidity 20 - 80% RH
Safety and EMC
compatibility
According to EC-directives; CSA (UL) approved
Installation class II
Pollution degree 2
External circuit
insulation
Basic (single layer of insulation)
Altitude Up to 2000 m
Dimensions Without cooling: 300 mm (W) x 510 mm (D) x 360
mm (H)
With cooling: 300 mm (W) x 575 mm (D) x 360
mm (H)
Weight Without cooling: 19 kg
With cooling: 21 kg
Maximum weight that
can be placed on the
instrument
65 kg
Electrical requirements 95 - 240 Volt AC ± 10%; 50 - 60 Hz; 200 VA
Viscosity range 0.1 - 5 cP
Sampling
Sample capacity 2 microtiter plates according to SBS standards; 96-well
high/low and 384-well low formats, 48- or 12-vial trays;
any combination of plates is allowed, except for 384
Low left and 96 High right.
Vial/plate
dimensions
Including septa or capmat: Maximum 47 mm height