Outlining general specifications C-3
Loop volume 1 - 5000 µL programmable, 10 mL loop optional
Dispenser syringe 500 µL standard or 2500 µL for prep option
Vial detection Missing vial/well plate detection by sensor
Headspace
pressure
Built-in compressor, but only for vials with septa
Switching time
injection valve
Electrically < 100 msec
Piercing precision
needle
± 0.6mm
Wash solvent Integrated wash solvent bottle
Wetted parts in
flow path
SS316, PTFE, Tefzel, glass, PEEK
Injection cycle time < 60 sec in all injection modes for 1 injection 100 µL
including 300 µL wash
Analytical performance
Injection modes Full loop, partial loop needle overfill, partial loop,
pressure-assisted sample aspiration
Reproducibility
(valid at 1.0 cP)
RSD 0.3% for full loop injections
RSD 0.5% for partial loop needle overfill injections,
injection volumes > 10 µL
RSD 1.0% for partial loop injections, injection
volumes > 10 µL
Carryover < 0.05% with programmable needle wash
Programming
Injection methods Full loop, partial loop needle overfill, and partial loop
Injection volume 0 µL - 10000 µL (with 1 µL increment), depending on
system settings
Maximum injection
volume
Full loop = loop volume
Partial loop needle overfill = ½ × of loop volume
Partial loop = (loop volume - [3 × needle volume])/2
Analysis time Maximum: 9 hr, 59 min, 59 sec
Sampling