1-6 Understanding Operating Principles
Exploring injection principles
The autosampler has three injection modes:
Full loop– For maximum injection volume precision
Partial loop needle overfill – For maximum flexibility
Partial loop – For zero sample loss
Defining pressure-assisted sample aspiration
For all injection modes, the instrument method can enable loop injection with
pressure-assisted sample aspiration. This option pairs an on-board
compressor with intelligent valve switching and highly accurate syringe
control, resulting in these benefits:
Helps aspirate viscous samples
Reduces the risk of bubbles in the sample line
Prevents unwanted sample movement in the needle
Pressurizing vial
Buffer tubing
HPLC pump
Sample loop
HPLC column
Air pressure
Syringe