Waters APGC - Removing the APGC source enclosure from the mass spectrometer

Default Icon
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-20 Setting up the APGC
Removing the APGC source enclosure from the mass
spectrometer
Required materials
Chemical-resistant, powder-free gloves
To remove the APGC source enclosure
:
1. Slide open the source’s interface door.
2. Disconnect the source interface cable from the connector on the
instrument.
3. Using two hands, grasp the source enclosure, and lift it vertically off the
two supporting studs on the source adaptor housing.
Warning: To avoid personal contamination with biologically
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when handling
source components.
Warning: To avoid burn injuries, take great care while
working with the source components, which can be hot.