Preparing the mass spectrometer for operation with the APGC 2-9
Installing the APGC ion chamber on the mass spectrometers ion
block
Required materials
Chemical-resistant, powder-free gloves
Combined, 2.5-mm Allen wrench and cone extraction tool
Dentist’s mirror
Requirement: During APGC operation with the Xevo TQD or SQ Detector 2
mass spectrometer, you must use the reduced-aperture, 0.36-mm, sample cone
supplied with the APGC, to avoid problems associated with unstable signal
intensity.
To install the APGC ion chamber:
1. Select the correct sample cone for the instrument you intend to use with
the APGC:
For Xevo TQD or SQ Detector 2 instruments, use the 0.36-mm
sample cone provided with the APGC. The cone is labelled “0.36mm”
on its base and side wall.
For all other instruments, use the sample cone that you retained in
the previous procedure (page 2-8, step 8)
2. Ensure that the O-ring is in position on the sample cone.
Warning: To avoid personal contamination with biologically
hazardous, toxic, or corrosive materials, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when installing
the ion chamber.
Warning: To avoid burn injuries, take great care while
working with the source components, which can be hot.