EasyManua.ls Logo

Wavecom TnP-500X - Page 23

Wavecom TnP-500X
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Wavecom Instruments Fully Integrated Test N Print Unit - User Manual
23
TnP500, TnP500W, TnP500X & TnP500WX all are required to be connected as per bellow. Use adapter cable that is
provided with X units 20A for 500X and 500WX units.
Diagram 1 For Testing Portable RCD Diagram 2 For Testing Fixed RCD -All models
For testing Portable RCD in TnPB series with 4.3 Motherboard and Internal Lithium-ion Battery [Separate Isolation
Transformer not required]
Note* Portable RCD must have hard re-set button (unlike RCDs like
in Diagram1 above where re-set button only works when RCD is
powered. In such cases, please connect TnPB series as per diagram
1)
Connect Portable RCD with hard re-set switch- as per diagram 3.
Enable Portable RCD from Main Menu B > F3 / RCD Tests > F1 /Time
Test > F3 /Settings > F3 / Change RCD Type > select F2 or F3 > Select
F3 to enable PRCD or Enter to disable PRCD.
Enable Portable RCD from Main Menu B > F3 / RCD Tests > F2 /
Ramp Tests > > F2 / Change RCD Type > select F2 or F3 > Select sF3
to enable PRCD or Enter to disable PRCD.
For testing Fixed / Wall/ Board RCD in TnPB series; Power
the unit from RCD board / GPO, connect as per diagram
2 in previous page. Disable PRCD function as per the flowchart above, to test fixed RCD / wall / Board RCD.
To do RCD test, select F3 in Menu B, press Return/Enter button.
To enter the RCD menu from Main Menu B, press and release the F3 button.
Diagram 3PRCD setup for TnPB series

Related product manuals