Technical Data 31
13.1 Safe distances
The ACCUVAC Rescue is intended for operation in an electromagnetic environment in which high-frequency
interference variables are controlled. The customer or user of the ACCUVAC Rescue can help avoid electro-
magnetic interference by maintaining the minimum safe distance between portable and mobile high-frequen-
cy telecommunication devices (transmitters) and the ACCUVAC Rescue depending on the rated output of the
transmitter as given below.
Recommended safe distances between portable or mobile HF telecommunication
devices (e.g. mobile phones) and the ACCUVAC Rescue
Rated output of
HF device
in W
Safe distance depending on transmission frequency
in m
150 kHz - 80 MHz
d=(3,5/V1) x
P
80 MHz - 800 MHz
d=(3,5/V1) x
P
800 MHz – 2,5 GHz
d=(3,5/V1) x
P
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33