Maintenance 9
6. Maintenance
The ACCUVAC Rescue needs no maintenance,
but please be sure to observe the intervals speci-
fied for regular final checks (see ”7.1 Intervals“ on
page 9).
To maintain battery operation and service life we
recommend performing calibration every 8 weeks
according to item 7.1.2 of the description and
operating instructions. This process involves the
necessary specific battery discharging and
recharging.
We recommend that you have any servicing, such
as inspections and repair work, carried out by the
manufacturer – Weinmann – or by expert
personnel.
6.1 Disposal
Do not dispose of the unit with domestic waste. For proper waste disposal of the equipment, please
contact an approved and certified waste disposal site for electronic goods. Ask your Environmental
Officer or town council for the address.
7. Function checks
If the final check reveals defects or deviations from
the specified values, the ACCUVAC Rescue must
not be used again until the faults have been recti-
fied.
We recommend that you always keep a stock of
the following:
Aspiration tube 27, WM 10662
Nozzle with fingertip 26, WM 10666
Filter 20, WM 10675
7.1 Intervals
To ensure that a properly functioning
ACCUVAC Rescue is always available, it is essen-
tial to observe the following intervals.
Before every use
Perform a function check (see ”7.2 Performing
the function check“ on page 10).
After every use
Clean, disinfect and/or sterilize the unit and
its parts (see operating instructions “5.
Cleaning and disinfecting”);
Perform a function check (see ”7.2 Performing
the function check“ on page 10).
Every 6 weeks
Check the power pack charge level by switch-
ing on the ACCUVAC Rescue and reading the
capacity indicator. If the capacity is 50% or
less, you should recharge the power pack (see
operating instructions “4.4 Charging the
ACCUVAC Rescue”).
At least every 6 months
Perform a function check (see ”7.2 Performing
the function check“ on page 10).
Make a visual inspection of the muffler for
clogging. If it is clogged, fit a new muffler (see
”9.13 Changing the muffler“ on page 25).
After all repairs
Clean, disinfect and/or sterilize the unit and
its parts (see operating instructions “5. Clean-
ing and disinfecting”);
Perform a function check (see ”7.2 Performing
the function check“ on page 10).