46 Chapter 5 Performing ECG Tests Welch Allyn CP 100 Electrocardiograph
Figure 28. Electrode Placement Locations — Augmented Pediatric
Electrodes
AHA IEC Locations
A V1
red
C1
red
Fourth intercostal space at right sternal border.
B V2
yellow
C2
yellow
Fourth intercostal space at left sternal border.
C V4
blue
C4
brown
Fifth intercostal space at left of midclavicular line.
D V7
orange
C7
black
Posterior axillary line at the same horizontal level as V4.
E V6
purple
C6
purple
Mid-axillary line on same horizontal level as V4.
F LA
black
L
yellow
Just above left wrist on inside of arm.
G LL
red
F
green
Just above left ankle.
H RL
green
N
black
Just above right ankle.
I RA
white
R
red
Just above right wrist on inside of arm.
J V3R
green
C3R
green
Mirror position to standard V3 placement with respect to the mid-
sternal line.
I
A
H
B
J
C
E
D
F
G
To use the augmented pediatric lead set, you
must enable it in the settings. See “Turning the
Augmented Pediatric Lead Set On and Off” on
page 41.