EasyManua.ls Logo

Wheaton W375030-A - 11.0 Making Batch and Single Dose Dispenses; 11.1 Flow Rate

Default Icon
35 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WHEATON
®
OmniSpense
®
ELITE Peristaltic Pump
WHEATON | 800.225.1437 (U.S. & Canada Only) | 856.825.1100 | www.wheaton.com | P/N 50099067 Rev. 5/15/12 WMAN_001
15
Press the START / STOP button to end the calibration dispense.
Use the UP and DOWN buttons and the LEFT and RIGHT buttons to enter the measured volume:
Press OK to enter the volume and complete the calibration procedure. Press OK again, to return to the
HOME screen.
11.0 Making Batch and Single Dose Dispenses
From the HOME screen, press OK to navigate to the DISPENSE SETUP screen:
11.1 Flow Rate
Use the UP and DOWN and the LEFT and RIGHT keys to enter the desired flow rate for the dispense.
Press OK when done.
Unit is in Dispense
Setup Mode
Active Keys
Maximum attainable flow rate based on tubing size, number of heads and calibration

Table of Contents

Related product manuals