EasyManua.ls Logo

WLKATA Mirobot - Appendix I: Calibration Operation; Calibration Procedure Steps

WLKATA Mirobot
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WLKATA Mirobot Manipulator User Manual
Document version 1.0 User Manual Copyright© Beijing Tsinew Technologies Co., Ltd.
Appendix I
WLKATA Mirobot Calibration Operation
Calibration Operation Procedure:
The Mirobot is calibrated before leaving the factory. In each time the manipulator finishes
the HOMING action, the joint 1, joint 2, joint 3, joint 4 and joint 5 should be homed to the
pre-designed homing position.
After the HOMING action, if one or more of the joints is/are not in the pre-designed homing
position, one need to re-calibrate the manipulator. The Calibration procedure is explained as
below:
Step 1: Turn on the power of the robotic arm. Open the "COMMAND" interface of Mirobot
Studio software, enter "M50" and click "Send" to unlock each axis.
Step 2: Enter the "SETTING" interface, click "Calibration-Start".
Open Mirobot Studio settings
01
Start calibration
02
Calibrate the position of each axis
03
Complete calibration
04
Home Mirobot
05

Table of Contents