V5 series inverter 
77 
P7.06  Min setting value  Range: 0.0~Max setting value  0.0% 
P7.07  Feedback value corresponding to min setting    Range: 0.0~100.0 (%)  0.0% 
P7.08  Max setting value  Range: min reference~100.0 (%)  100.0% 
P7.09  Feedback value corresponding to max setting  Range: 0.0%~100.0 (%)  100.0% 
P7.06~P7.09 defines the relationship curve of analog close-loop setting value and expected feedback. See Fig4-36. 
 
Fig. 4-36 setting and feedback curve 
P7.10  Proportional gain KP  Range: 0.000~9.999  0.050 
P7.11  Integral gain KI  Range: 0.000~9.999  0.050 
P7.12  Sampling cycle T    Range: 0.01~10.00s  1.00s 
The bigger the proportion gain KP, the quicker the response, but too big may lead to surge easily. 
Only adjust with proportiaon gian KP, deviation can not be eliminated, integral gain KI can be used to form PI control to 
eliminate the deviation. The bigger the Ki, the quicker the response of changing deviation, but too big of the Ki may lead to 
surge easily. 
Sampling cycle T is the sampling cycle of feedback value. PI regulator calculates once in every sampling cycle, the longer 
the cycle, the slower the response. 
P7.13  Limit of deviation  Range: 0.0~20.0(%)  2.0(%) 
The max deviation value of close-loop setting is shown in Fig.4-37, if the feedback value is in this range, PI regulator will stop adjusting. 
This function is useful to improve accuracy and stability of the system. 
 
        Fig.4-37 Limit of deviation            Fig.4-38 close-loop preset frequency running 
 
P7.14  Close loop adjustment characteristic  Range: 0, 1  0 
0: Positive. The motor speed increases with the increasing of setting value. 
1: Negative. The motor speed decreases with the increasing of setting value. 
Note: Define the relationship between setting value and speed. 
P7.15  Integral adjustment selection  Range: 0, 1  0 
0: Stop integral adjustment when the frequency reaches upper limit or lower limit. 
1: Continue the integral adjustment when the frequency reaches upper limit or lower limit. 
For the system need quick response, it is better to cancel the continue integral adjustment.