EasyManua.ls Logo

XS Instruments 60 VioLab Series - 7. Setup menu; pH SETTINGS; ORP SETTINGS; COND SETTINGS

Default Icon
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
Rev. 1.0 31/01/2020
EN
String
Meaning
The instrument is in measure mode.
The instrument is in calibration mode
(automatic or manual in
relation to the user's choice).
The user is in the setup mode.
The configuration menus can concern the characteristics of the
parameters or the general setting of the instrument.
The instrument is in the Recall Memory mode.
The data that have been stored are being displayed by running the
manual or automatic Data Logger.
Setup menu
In measure mode, press the key to enter SETUP mode, select the parameter you want to edit
by using the directional keys and confirming with .
pH 60 VioLab
COND 60 VioLab
PC 60 VioLab
pH SETTINGS COND SETTINGS pH SETTINGS
ORP SETTINGS TDS SETTINGS ORP SETTINGS
LOG SETTINGS LOG SETTINGS COND SETTINGS
SETTINGS SETTINGS TDS SETTINGS
LOG SETTINGS

Table of Contents

Related product manuals