Access this menu to select the information to print in the header of a Data Logger on a printer:
SIMPLE: -default setting- The device model, Serial Number and date and time of the last calibration
are printed in the header.
COMPLETE: In addition to the information of the SIMPLE format, the calibration data are also printed
(available in the "CALIBRATION DATA" setup menus).
Note: The header is printed in the following cases:
When a printout of values for a certain parameter is started for the first time
When a new calibration is performed
Example of automatic Data Logger mode
Example: automatic pH recording on internal memory every 2 minutes
Access the LOG SETTING P8.0 setup menu.
Enter the LOG TYPE P8.1 menu, press and move with directional keys to MINUTES.
Use the directional keys to change the number that flashes on the display. Enter "2" and confirm with
.
Go back to measure mode and go to the pH screen .
The icon is lit in the lower string of the display, which indicates that an automatic frequency Data
Logger has been set. Press to start recording; the icon starts flashing, indicating that the
storage is in progress. The number next to the M+ icon indicates how much data has been saved for
that parameter.
Press again to end the recording.
Note: automatic recording is suspended when the measurement parameter is changed.
Example of manual Data Logger mode
Example: recording a Conductivity value in manual mode
Access the LOG SETTING P8.0 setup menu.
Enter the LOG TYPE P8.1 menu, press key and move with the directional keys to MANUAL.
Confirm with and return to measure mode, go to the COND screen .
Press to save the value. The number next to the M+ icon indicates how much data has been saved
for that parameter.
Note: Manual or automatic saving of a value is confirmed by a sequence of flashes of the green LED
Example of printing a value
Example: manual printing of a pH value with calibration data too
Access the LOG SETTING P8.0 setup menu.
Enter the LOG TYPE P8.1 menu, press and move with the directional keys to MANUAL. Confirm
with .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the XS Instruments 60 VioLab Series and is the answer not in the manual?

XS Instruments 60 VioLab Series Specifications

General IconGeneral
BrandXS Instruments
Model60 VioLab Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals