EasyManua.ls Logo

XS Instruments 60 VioLab Series - ORP Parameter Setup; Composition of the setup menu for the ORP parameter

Default Icon
41 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
Rev. 1.0 31/01/2020
EN
ORP Parameter Setup
In measure mode press the key to access the SETUP menu.
Use the directional keys to move to ORP SETTINGS P2.0 and access the menu by pressing the key
.
Move with the keys and to select the program to access.
The table below shows the Setup menu structure for the ORP parameter; for each program there are the
options that the user can choose and the default value:
Composition of the setup menu for the ORP parameter
Program
Description
Options
Factory Default Settings
P2.6
CALIBRATION DATA
-
-
P2.7
SET DUE CAL
NO HOURS - DAYS
NO
P2.8
RESET SETTINGS
YES NO
NO
P2.9
TEMPERATURE CAL
YES NO
-
P2.6 Calibration Data
Access this menu to view (select “VIEW) or print (select “PRINT”) information on the last calibration
performed:
VIEW: The following screens will scroll automatically on the display:
First screen: CALIBRATION DATE and TIME.
Second screen: OFFSET value of the electrode expressed in mV.
Third screen: TEMPERATURE at which the calibration was performed.
PRINT: The following information are printed:
Model and Serial Number of the instrument.
Calibration Date and Time
OFFSET value expressed in mV
Calibration TEMPERATURE
IMPORTANT: Make sure that the printer is connected, turned on and that the paper roll and cartridge are correctly
positioned. For more information, refer to the printer user manual.
NOTE: Use the original printer indicated by the manufacturer only. For information, contact your local distributor.
P2.7 ORP calibration deadline
Access this menu to set a calibration deadline; this option is fundamental in GLP protocols.

Table of Contents

Related product manuals