HandyLab 100 Operation
29
ba77062d02 12/2017
4.4.5 pH calibration
Why calibrate? pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is dis-
Calibration / One
point calibration
Yes
No
Quick calibration with 1
buffer
Calibration / Calibra-
tion interval
1 ... 999 d Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring screen.
Calibration / Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the
Nernst slope of -59.2 mV/
pH (100 x determined
slope/Nernst slope).
Man. temperature -25 ... +130 °C Entry of the manually
determined temperature.
For measurements without
temperature sensor only.
Temperature unit °C
°F
Temperature unit,
degrees Celsius or
degrees Fahrenheit.
All temperature values are
displayed with the
selected unit.
Resolution pH 0.001
0.01
0.1
Resolution of the pH dis-
play:
Resolution mV 0.1
1
Resolution of the mV dis-
play:
Stability control On / Off Switches on or off the
automatic stability control
during measurement (see
section 4.3.3)
Reset - Resets all sensor settings
to the delivery condition
(see section 4.6.1).
Menu item Possible
setting
Explanation

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Xylem SI Analytics HandyLab 100 and is the answer not in the manual?

Xylem SI Analytics HandyLab 100 Specifications

General IconGeneral
BrandXylem
ModelSI Analytics HandyLab 100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals