Operation HandyLab 100
46
ba77062d02 12/2017
4.6 Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1 Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH The following settings for pH measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu
for calibration and measurement settings. To open the settings, display
the required parameter in the measured value display and press the
<F1>/[Menu] or <ENTER> key.
Setting Default settings
Buffer TEC
Cal. interval 7 d
Unit for slope mV/pH
Measured parameter pH
Resolution pH 0.001
Resolution mV 0.1
Asymmetry 0 mV
Slope -59.2 mV
Man. temperature 25 °C
One point calibration Off
Stability control On
Temperature unit °C