What to do if... HandyLab 100
52
ba77062d02 12/2017
Obviously incorrect
measured values
Meter does not react to
keystroke
You want to know which
software
version is in the meter
Cause Remedy
pH electrode:
pH electrode unsuitable Use suitable electrode
Temperature difference
between buffer and test sam-
ple too great
Adjust temperature of buf-
fer or sample solutions
Measurement procedure not
suitable
Follow special procedure
Cause Remedy
Operating condition unde-
fined or EMC load unallowed
Processor reset:
Press the <ENTER> and
<On/Off> key simultane-
ously
Cause Remedy
E.g., a question by the ser-
vice department
Switch on the meter.
Open the menu, <F1
_> /
[Menu] / Storage & config /
SystemService informa-
tion. The instrument data
are displayed.