Lab 855 pH value
ba77143e02 12/2017 19
Calibration determines and stores the current values of the zero point and
slope. Thus, you should calibrate at regular intervals.
5.2.2 When to calibrate?
When the calibration interval has expired
Routinely within the framework of the company quality assurance
After connecting another combination electrode
5.2.3 Automatic calibration (AutoCal)
Use any one to three buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with buffers (TEC) is described. When other buffer sets are
used, other nominal buffer values are displayed. Apart from that, the procedure
is identical.
1. Start the calibration with <CAL>.
The calibration display for the first buffer appears.
2. When measuring without temperature sensor: Temper the buffer or
measure the current temperature.
3. If necessary, press <CAL> to select the buffer set used ([AutoCal TEC],
[AutoCal DIN]).
Ct1 or Cd1 is displayed.
4. Thoroughly rinse the pH electrode with deionized water.
5. Immerse the pH electrode in the first buffer solution.
6. When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.
7. Start the measurement with <ENTER> .
The measured value is checked for stability (stability control).
The [AR] status display flashes.
The electrode voltage (mV) or the nominal value of the buffer is displayed
(setting: see section 7.1.1 C
HANGING THE SETTINGS FOR PH MEASURE-
MENTS, page 28).
8
4
2
1
T
C
°
C
pH
AutoCal TEC
TP