EasyManua.ls Logo

Xylem SI Analytics Lab 855 - Calibration Points; Calibration Data

Xylem SI Analytics Lab 855
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lab 855 pH value
ba77143e02 12/2017 23
14. When measuring without temperature sensor: Enter the temperature of
the buffer with <><>.
15. Start the measurement with <ENTER> .
The pH value of the buffer solution is displayed.
The measured value is checked for stability (stability control).
The [AR] status display flashes.
16. Wait for the measurement with stability control to be completed.
17. Set the nominal pH value of the buffer solution with <><>.
18. Accept the calibration value with <ENTER>.
The calibration record is displayed.
5.2.5 Calibration points
Depending on the number of buffer solutions used, the meter determines the
following values and calculates the calibration line:
5.2.6 Calibration data
Displaying the calibration data
1. Display the calibration data in the measured value display with <CAL__>.
The value for the asymmetry (ASY) is displayed.
Calibration
Values deter-
mined
Displayed calibration data
1-point Asy
Zero point = Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point Asy
Slo
Zero point = Asy
Slope = Slo
The calibration line goes through both cali-
bration points.
3-point Asy
Slo
Zero point = Asy
Slope = Slo
The calibration line is calculated by linear
regression.
You can display the slope in the units, mV/pH or % (see section
5.2.6 C
ALIBRATION DATA, page 23).

Table of Contents

Related product manuals