EasyManua.ls Logo

Xylem SI Analytics Lab 855 - Error Message OFL, UFL; Orp; No Stable Measured Value

Xylem SI Analytics Lab 855
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Lab 855 What to do if...
ba77143e02 12/2017 37
10.1.3 Error message OFL, UFL
The measured value is outside the measuring range.
The measured value is obviously incorrect.
10.2 ORP
10.2.1 No stable measured value
Cause Remedy
• The used buffer solutions do not
agree with the set buffer set
• Set different buffer set
or
• Use different buffer solutions
• Buffer solutions too old • Use only once.
Note the shelf life
• Buffer solutions depleted • Change solutions
Cause Remedy
• Cable broken • Replace the electrode
• Gel electrolyte dried out • Replace the electrode
• Air bubble in front of the junction • Remove the air bubble
• Air in the junction • Extract air or moisten the junction
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause Remedy
• Junction contaminated • Clean the junction
• Pt-ORP electrode contaminated • Clean the Pt-ORP electrode
Cause Remedy
• ORP value not stable • Measure with air excluded if neces-
sary
• Temperature not stable • Adjust temperature if necessary

Table of Contents