Operation Multi 3630/3620 IDS
10 ba77045d02 03/2020
4.1 Measuring modes while operating
the OxiTop
®
-IDS measuring heads
Red Measurement is running.
The measuring head is in a critical condition:
The battery is nearly discharged
or
The pressure in the sample bottles is too high.
Suitable actions, see operating manual of your measuring
head.
Green The measurement is finished.
The status LED flashes green for approx. 24 hours.
The measuring head OxiTop
®
-IDS can be reached via
radio.
Measuring mode Explanation
BOD Measurement of up to 12 parallel sample
processes per overall sample for a BOD
x
(x
= 0.5h ... 180d)
Automatic sample statistics with averaging
The AutoTemp function can be switched on
The GLP function can be switched on
BOD stan-
dard
The measuring range and sample volume
can be selected from seven combinations.
The incubation temperature is 20 °C
BOD special Measuring range, Dilution 1 +, Sample vol-
ume and Bottle volume can be freely
adjusted
The incubation temperature (5 °C to 40 °C)
can be set up
Pres-
sure p
Pure pressure measurement
Measuring range:
– 500 ... 1250 hPa
(OxiTop
®
i/OxiTop
®
-IDS)
– 500 ... 1500 hPa
(OxiTop
®
-IDS /B)
A Alarm diff. pressure can be set
Up to 10 manually triggered measurements
(momentary values) are saved with pres-
sure value and point in time
The GLP function can be switched on