Operation Multi 3630/3620 IDS
16 ba77045d02 03/2020
4.3.2 Measurement settings of the sample
The measurement settings for the sample are defined when the sample is reg-
istered within a set.
Default settings are printed in bold.
Number of measured
values
Standard
Max.
1 measured value per day
360 measured values for the specified mea-
suring duration
LED on
off
Switching on/off the LED indication of the
operating statuses (at the OxiTop
®
-IDS
measuring head).
The LED (at the OxiTop
®
-IDS measuring
head) can be switched off to save energy.
Eco mode (radio) on
off
Energy saving function to switch on or off the
OxiTop
®
-IDS measuring heads.
Only for Eco mode
(radio) = on:
Radio connection
from
Radio connection
until
each
0.00 ... 23.00
Entry of the time (hours, begin and end) at
which the measuring heads can be reached
via radio every day (LED magenta).
Outside the specified times, the radio con-
nection of the measuring heads is switched
off in order to save energy (LED blue).
The restricted radio connection is only active
during a measurement.
Menu item Possible setting Explanation
The general settings for the set apply to the entire meter. All other
settings always apply to all OxiTop
®
-IDS measuring heads in the
current set.
Menu item Possible setting Explanation
Sample name max. 8 characters The 8 characters can be freely
selected.
Preselected is an A.
An index is also assigned automati-
cally.
Measuring range
(for BOD standard)
0 ... 40 / 80 / 200 mg/l
0 ... 400 mg/l
0 ... 800 / 2000 / 4000 mg/l
The measuring range expected for the
sample can be selected from 7 perma-
nent ranges.
Selecting the measuring range defines
the Sample volume. The correspond-
ing Sample volume is displayed as
Info.
Calculation
(for BOD special)
Measuring range
Sample volume
Bottle volume
Parameter to be calculated
The parameters not calculated are set
in the Settings menu.