EasyManua.ls Logo

Yamato UDS-1V-WP - Page 37

Yamato UDS-1V-WP
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
35
Display Cause Troubleshooting
The initial error (A)
Turned on the scale while
a heavy item was placed.
Remove the item from the pan, and
then press
.
The initial error (B)
Turn on the scale while a
foreign material was stuck
between the main body
and the pan.
Check if a foreign material is not stuck
or does not contact the pan. Then
press
. When the scale does
not return to normal, contact the
retailer from which the scale was
purchased for troubleshooting.
The initial error (C)
Pressed
while a
heavy item was on the
pan in the middle of the
operation.
The weight of the item is beyond the
level where t
is operative.
Remove the item from the pan, and
then press
.
The initial error (D)
Turned on the scale while
the item was left on a pan.
And then, removed it and
pressed
.
Make sure that nothing is left on the
pan and turn on the scale again.
Trouble in electronic circuit
This message appears
when a trouble was
detected in the electronic
circuit.
Turn off the scale. And reapply the
power after a brief interval.
When the scale does not return to
normal, contact the retailer from which
the scale is purchased for
troubleshooting.
Trouble in control unit
This message appears
when a strong static
electricity was applied or
any other trouble was
detected in the electronic
circuit.
Turn off the scale. And reapply the
power after a brief interval.
When the scale does not return to
normal, contact the retailer from which
the scale was purchased for
troubleshooting.

Related product manuals