EasyManuals Logo

YL Instruments YL9100 HPLC User Manual

YL Instruments YL9100 HPLC
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #112 background imageLoading...
Page #112 background image
YL9112 Isocratic Pump
Problem Cause How to fix
Pressure upper
limit is loaded.
-Tubing inside is blocked.
-Check valve is blocked.
-Solvent is changed.
-Replace blocked tubing by
loosening to be from tail side
in turn.
-Replace outlet check valve.
-Wait until solvent is completely
changed.
Pressure
increases or
decreases.
-Change of solvent is
incomplete.
-Wait until change is completely
performed.
-Wait until pressure is stable.
Solvent is not
flowed out
-Air bubble is in pump head.
-Air separation status of
solvent is bad.
-Check valve is not good.
-Liquid containing oil flows
into head part.
-Perform prime/purge again.
-Take measures so that air
separation condition of solvent
may be good, and perform
prime/purge again.
-Wash or replace check valve.
-Remove oil in head sufficiently
with strong organic solvent
using prime/purge port.
Pressure is
unstable.
-Check valve is bad or
defected.
-Air separation or mixing
conditions of solvent is
bad.
-Compression compensating
reference
value of configuration
mode is wrong.
-Cam shaft is loosened.
-High pressure seal was
worn.
-Pump head was loosened.
-Wash or replace check valve.
-Use mobile phase with well mixed
solvent, and improve air separation
condition.
-Initialize compensation value or
re-input proper value.
-Tighten wrench bolt of cam shaft
-Replace high pressure seal.
-Tighten nut of pump head.
30

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YL Instruments YL9100 HPLC and is the answer not in the manual?

YL Instruments YL9100 HPLC Specifications

General IconGeneral
BrandYL Instruments
ModelYL9100 HPLC
CategoryLaboratory Equipment
LanguageEnglish