YL9120 UV/Vis Detector
14) Path Length : 10 mm ( Analytical cell )
15) Cell Volume : 10 uL ( Analytical cell )
16) Pressure limit : 1500psi ( Analytical cell )
17) Wetted Materials : 316 stainless steel, Quartz, Teflon
18) Analog Outputs
A. Two, software selectable : Absorbance Energy, Sample Energy, Reference Energy
19) Communications : LAN
20) Power-up Diagnostics : Optics and electronic diagnostic routine
21) Power-up Wavelength Verification : Automatic on power up via internal holmium filter and D2
lamp
22) 5-Point Wavelength Calibration : On demand via internal holmium filter and D2 lamp
23) Safety & maintenance : Leak detection, Diagnostics, Error detection
24) Dimensions : 385 X 160 X 565mm (width X height X depth)
25) Line Voltage : 110 or 220 VAC, ±10%
26) Line frequency : 50/60Hz, ±5%
27) Power consumption : 100W
2