YL9111 Binary Pump
L9111
[Fig. 6] Connection between YL9111 Binary pump and YL9100 HPLC system
Rinse port tubing connects on the each pump head using 1/4-28 fitting as same as [Fig. 7], and fill the
20~50% Methanol. The rinse pump inside of YL9111 Binary pump circulates the rinse solvent into the
pump head every minute. Check and replace rinse solvent once a week at least.
[Fig. 7] Tubing connection of rinse port
11