3-22
IM AQ6370E-02EN
7.
Tap Cancel during wavelength calibration to cancel the wavelength calibration process.
Note
Always perform wavelength calibration after turning ON the power to the instrument and allowing the
warm-up to finish.
If you are using this instrument for the first time or using it after it has been severely shaken while being
moved, you must perform the alignment adjustment procedure after warm-up ends.
If the wavelength error of the instrument is outside of ±5 nm, you cannot perform wavelength calibration
with the internal reference light source.
(Readjustment required. Contact your nearest Yokogawa dealer.)
3.6 Wavelength Calibration