5-12
IM AQ6370E-02EN
No. Message Cause of Warning
234 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
235 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
236 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
237 Monochromator error Entered emulation mode because an abnormality with the operation
of the monochromator occurred.
No. 300 - 399: Errors during the execution of program functions
300 Parameter out of range A variable value is out of range or is not defined for a command that
sets a parameter using variables.
302 Scale unit mismatch There is a difference between the Y-axis scale of the active trace
and the unit of a parameter in the “LINE MKR 3 or 4” command.
303 No data in Active trace Setting of the moving marker, a peak (or bottom) search, or
activation of the analysis function was made with no data in the
active trace.
304 Marker value out of range Specified wavelength was out of the sweep range in the moving
marker or line wavelength marker setting command.
305 No data in trace A or B No waveform data in traces A or B when executing the “EDFA NF”
command.
306 Invalid data Trace had no data when attempting to save it to memory or to write
it to FD/HDD.
307 Unsuitable Write item All data items were OFF at execution of “WRITE DATA”.
320 Undefined variable A command containing an undefined variable was executed.
321 Variable unit mismatch The unit of each variable does not agree within a command
containing two or more variables.
322 Overflow An overflow occurred in an arithmetic operation.
323 Undefined marker variable A command containing a marker-value variable was executed when
no marker had been displayed.
324 Invalid marker variable A command containing the corresponding variable was executed at
a time other than immediately after execution of a spectrum width
search, peak search, etc.
325 Undefined line number GOTO command’s jumping destination is a number other than 1 to
200.
326 F1 greater than F2 F1>F2whenthe“IFF1≤@@@@@≤F2”commandwas
executed.
345 Option does not respond No response from an external device.
346 Option is not connected No external device is connected.
360 Disk full No file can be created due to insufficient free space on the USB
storage medium.
361 USB Storage not inserted USB storage medium not inserted.
362 USB Storage is write protected The USB storage device is write protected.
363 USB Storage not initialized USB storage not initialized. Or, it has been formatted in a format not
supported by this instrument.
364 Directory full Directory is full, therefore no file can be created.
365 File not found The specified file cannot be read because it has not been found. Or,
the file does not exist on the disk.
366 File is write protected The file is specified to be read only, so that it cannot be rewritten or
deleted.
367 No data No data to store.
368 File is not a trace file A file cannot be read because it is not a trace file.
369 Illegal file name A file cannot be saved due to an incorrect file name.
370 File type mismatch Loading or saving the file is not possible because the specified type
of file does not match that of the command.
380 Undefined program An attempt was made to run a program that is not defined.
381 Syntax error Command incorrect (a program has been rewritten for some
reason).
5.7 Warning Display Function