24
IM CA500-02EN
Installation Conditions
Install the instrument in a place that meets the following conditions.
Operating Altitude and Ambient Temperature and Humidity
Use the instrument in the following environment.
Ambient temperature -10°C to 50°C
Ambient humidity 80% RH or less for -10°C to 40°C, 50% RH or less
for over 40°C
No condensation
Operating altitude Up to 2000 m
Note
To ensure high accuracy, operate the instrument within 23°C ± 5°C.
When using the instrument in a place where the ambient humidity is 30% or less, take
measures to prevent static electricity such as using an anti-static mat.
Condensation may occur if the instrument is moved to another place where the ambient
temperature or humidity is higher, or if the temperature changes rapidly.
In such cases, allow the instrument adjust to the new environment for at least an hour
before using the instrument.
Install the instrument in a proper location by referring to “Safety Precautions” on
page 8.
Do not install the instrument in the following places, as it can cause incorrect
measurements or damage the instrument.
In direct sunlight or near heat sources
In an environment with excessive amounts of soot, steam, dust, or corrosive gas
Near strong magnetic field sources
Near noise sources, such as high-voltage equipment or power lines
Where the level of mechanical vibration is high
On an unstable surface
Outdoors or in locations subject to rain or water
2 Preparation before Use