EasyManua.ls Logo

YOKOGAWA CA550 - Chapter 3 Measurement; 3.1 DC Voltage Measurement; 3.2 DC Current Measurement

YOKOGAWA CA550
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
IM CA500-02EN
5.4 Temperature Measurement (RTD) and
Resistance Source for RTD Calibration
CA500
RTD
Coef Source/measurement accuracy (1 year,
t: source/measurement temperature)
Allowable
Excitation
Current
Standard,
Citation
Temperature Range
CA500
±
[°C]
PT100
3851 -200.0°C ≤ t < +100.0°C 0.3 0.1 to 3 mA IEC60751
1
+100.0°C ≤ t ≤ +800.0°C 0.3+(t-100) x
0.033%
3850 -200.0°C ≤ t < +100.0°C 0.3 0.1 to 3 mA JIS C 1604
1989
(Pt100)
+100.0°C ≤ t ≤ +630.0°C 0.3+(t-100) x
0.033%
3916 -200.0°C ≤ t < +100.0°C 0.3 0.1 to 3 mA JIS C 1604
1989
(JPt100)
+100.0°C ≤ t ≤ +510.0°C 0.3+(t-100) x
0.033%
3926 -200.0°C ≤ t < +100.0°C 0.3 0.1 to 3 mA Minco
Application Aid
#18
+100.0°C ≤ t ≤ +630.0°C 0.3+(t-100) x
0.033%
PT200
3851 -200.0°C ≤ t < +100.0°C 0.3 0.05 to
3 mA
IEC60751
1
+100.0°C ≤ t ≤ +630.0°C 0.3+(t-100) x
0.033%
PT500
3851 -200.0°C ≤ t < +100.0°C 0.3 0.05 to
0.6 mA
IEC60751
1
+100.0°C ≤ t ≤ +630.0°C 0.3+(t-100) x
0.033%
PT1000
3851 -200.0°C ≤ t < +100.0°C 0.2 0.05 to
0.6 mA
IEC60751
1
+100.0°C ≤ t ≤ +630.0°C 0.2+(t-100) x
0.033%
Cu10
427 -100.0°C ≤ t ≤ +260.0°C 1.5 0.1 to 3 mA Minco
Application Aid
#18
Ni120
627 -80.0°C ≤ t ≤ +260.0°C 0.2 0.1 to 3 mA Minco
Application Aid
#18
PT50
3851 -200.0°C ≤ t < +100.0°C 0.4 0.1 to 3 mA IEC60751
1
+100.0°C ≤ t ≤ +630.0°C 0.4+(t-100) x
0.033%
PT50G
- -200.0°C ≤ t < +100.0°C 0.4 0.1 to 3 mA GOST R
8.625-2006
+100.0°C ≤ t ≤ +800.0°C 0.4+(t-100) x
0.033%
PT100G
- -200.0°C ≤ t < +100.0°C 0.3 0.1 to 3 mA GOST R
8.625-2006
+100.0°C ≤ t ≤ +630.0°C 0.3+(t-100) x
0.033%
Cu50M
- -180.0°C ≤ t ≤ +200.0°C 0.4 0.1 to 3 mA GOST R
8.625-2006
Cu100M
- -180.0°C ≤ t ≤ +200.0°C 0.3 0.1 to 3 mA GOST R
8.625-2006
5. Specifications