EasyManua.ls Logo

YOKOGAWA CA550 - 2.7 Setting the 0% and 100% Values

YOKOGAWA CA550
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
IM CA500-02EN
45
5.2 DC Voltage, DC Current, Resistance and Pulse
Measurement
DC Voltage Measurement
Range Resolution
Measurement
Range
Accuracy (1 year)
±(% of reading + offset)
Notes
CA500 CA550
100 mV 1 μV ±110.000 mV 0.015% + 10 μV 0.015% + 5 μV Input resistance:
1 GΩ or more
5 V 0.1 mV ±6.0000 V 0.015% + 0.5 mV Input resistance:
Approx.1 MΩ
50 V 1 mV ±55.000 V 0.015% + 5 mV Input resistance:
Approx.1 MΩ
Accuracy guaranteed under +23°C±5°C, 20 to 80%RH
Add a temperature coefficient of 0.005% of Range/°C for -10°C to +18°C and +28°C
to +50°C.
DC Current Measurement
Range Resolution
Measurement
Range
Accuracy (1 year)
±(% of reading + offset)
Notes
CA500 CA550
50 mA 1 μA ±60.000 mA 0.015% + 3 μA 0.010% + 2 μA Input resistance:
10 Ω or less
Accuracy guaranteed under +23°C±5°C, 20 to 80%RH
Add a temperature coefficient of 0.005% of Range/°C for -10°C to +18°C and +28°C
to +50°C.
5. Specifications

Other manuals for YOKOGAWA CA550

Related product manuals