IM CA500-02EN
41
4.2 Error Codes, Error Messages, and Corrective
Actions
Information and Warning Messages
No. Message Cause Corrective Action
01 Memory error Insufficient free internal
memory space
Delete unneeded data.
02 Low power error Low battery level
Low USB power
Replace the batteries, or switch to
a USB power supply that meets the
power supply specifications.
Error Messages
No. Message Cause Corrective Action
31 Memory error The measurement adjustment data saved
in EEPROM is inappropriate.
Servicing is required.
32 Memory error The source adjustment data saved in
EEPROM is inappropriate.
Servicing is required.
Communication Errors
No. Message Cause Corrective Action
11 Communication
Errors
Received an undefined
command
Send a correct command.
12 Communication
Errors
Incorrect command
parameter
Specify the parameter correctly.
13 Communication
Errors
Received a command that
cannot be executed in the
instrument’s current condition
Check the instrument condition, and
send the command again.
Self-Test Errors
No. Message Cause Corrective Action
53 RTC error RTC backup power supply
error
Initialize the date and time.
Servicing is required
81 CPU board error Measurement CPU error Servicing is required.
83 CPU board error Source CPU error Servicing is required.
4 Troubleshooting, Maintenance, and Inspection

Other manuals for YOKOGAWA CA550

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA CA550 and is the answer not in the manual?

YOKOGAWA CA550 Specifications

General IconGeneral
BrandYOKOGAWA
ModelCA550
CategoryTest Equipment
LanguageEnglish

Related product manuals