EasyManua.ls Logo

YOKOGAWA CA550 - Page 61

YOKOGAWA CA550
69 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
58
IM CA500-02EN
Thermocouple
Accuracy (1 year, t: measurement temperature)
Specification
Temperature Range Measurement
Accuracy
±
[°C]
S
-20.0°C ≤ t < 0.0°C 1.8 IEC60584-1
1, 2
0.0°C ≤ t < +100.0°C 1.2
+100.0°C ≤ t ≤ +1768.0°C 1.2
B
+600.0°C ≤ t < +800.0°C 1.3 IEC60584-1
1, 2
+800.0°C ≤ t < +1000.0°C 1.0
+1000.0°C ≤ t ≤ +1820.0°C 0.9
C
0.0°C ≤ t < +1000.0°C 0.6 IEC60584-1
1
+1000.0°C ≤ t ≤ +2315.0°C 0.6+(t-1000) x 0.06%
XK
-200.0°C ≤ t < 0.0°C 0.2+|t| x 0.10% GOST R
8.585-2001
0.0°C ≤ t < +300.0°C 0.2
+300.0°C ≤ t ≤ +800.0°C 0.3
A
0.0°C ≤ t < +1000.0°C 0.8 IEC60584-1
+1000.0°C ≤ t ≤ +2500.0°C 0.8+(t-1000) x 0.06%
D
(W3Re/W25Re)
0.0°C ≤ t < +300.0°C 1.6 ASTM E1751/
E1751M
+300.0°C ≤ t < +1500.0°C 1.0
+1500.0°C ≤ t ≤ +2315.0°C 2.0
G
(W/W26Re)
+100.0°C ≤ t < +300.0°C 1.6 ASTM E1751/
E1751M
+300.0°C ≤ t < +1500.0°C 1.0
+1500.0°C ≤ t ≤ +2315.0°C 2.0
PLATINEL II
0.0°C ≤ t < +100.0°C 1.6 ASTM E1751/
E1751M
+100.0°C ≤ t < +1000.0°C 1.6
+1000.0°C ≤ t ≤ +1395.0°C 2.0
PR20-40
0.0°C ≤ t < +500.0°C 11.0 ASTM E1751
+500.0°C ≤ t < +1000.0°C 4.0
+1000.0°C ≤ t ≤ +1888.0°C 2.0
Accuracy guaranteed under +23°C±5°C, 20 to 80%RH
Add a temperature coefficient of 0.05°C/°C for -10°C to +18°C and +28°C to
+50°C.
Display resolution of source/measured values: 0.1°C
Accuracy when using the included binding post (99045).
1 Complies also with JIS C 1602
2 The setting can be changed to comply with IPTS-68 (JIS C 1602 1981).
5. Specifications

Other manuals for YOKOGAWA CA550