3 - 2
SM 700820-01E
3.3 Equipment Required
Table 3.1 Equipment required
Equipment Critical Specification Recommended
DC Volt Standard Output Level : >1V YOKOGAWA 7651
Accuracy : <0.02%
Digital Volt Meter Accuracy : <0.01% YOKOGAWA 7551
Function Generator Frequency : 10 kHz YOKOGAWA FG120
Output Level : >12 Vp-p
Frequency Counter Frequency : >100 MHz YOKOGAWA TC110
Accuracy : < 0.005%
BNC-BNC Cable YOKOGAWA 3669 24
Clip-BNC Cable YOKOGAWA 3669 26
50 ohm Terminators Feed Thru Type YOKOGAWA 7009 76
NOTE The values shown in the specification column are those set in this service manual.
These values do not indicate the performances of the recommended equipment and
tools. Therefore, non-designated equipment and tools, which satisfy the specifications,
may be permitted for use.
The Function Generator setting values described in this service manual are those output
when 50 ohm load resistance is applied.
3.4 Dip Switch Setting on the CPU Board
Procedure
1) Remove the main unit cover and the two fixing plates.
2) Set the switch on the CPU Board (B9929LN/B9929LP) as following condition.
bit 1: OFF (not used)
bit 2: OFF (not used)
bit 3: OFF (not used)
bit 4: ON English Help Message
OFF Japanese Help Message
3.5 Rotary Switch Setting on the Printer Board
Procedure
1) Remove the main unit cover and the two fixing plates.
2) Set the switch on the Printer Board (B9919LF) corresponding to the Head Rank of the
attached Printer unit.
Rank A: A
Rank B: B
Rank C: C
Rank D: D
Rank E: E
Rank F: F
Rank G: 6
Rank H: 7