PERFORMANCE TEST
2 - 1
SM 700820-01E
2
Chapter 2 PERFORMANCE TEST
This chapter describes the following tests.
2.1 Introduction
2.2 Test Environment
2.3 Equipment Required
2.4 Self-diagnosis
2.5 Vertical axis DC voltage Accuracy Test
2.6 Frequency Response Test
2.7 Time-base Accuracy Test
2.8 Trigger Sensitivity Test
2.9 Trigger Accuracy Test
2.10 Logic input function Test
2.1 Introduction
The aim of the test is to check basic performance of the instrument. The order of the test
is just for convenience and it does not have any significant meaning. Please use recom-
mended equipment or their equivalents.
2.2 Test Environment
1) Please operate the instrument in the following environment.
• Ambient Temperature: 23±2 ˚C
• Humidity: 55±10 % RH
•Voltage of power supply: Specified Voltage ±1 %
• Frequency of power supply: Specified Frequency ±1 %
2) Warm up time
• More than 30 minutes after energizing the instrument.
•Confirm that self calibration is correctly executed after 30 minutes warm up.
(Please pay attention to warm up time of the each equipment which will be used in the
test.)