ADJUSTMENTS
SM 700820-01E
3
Figure 3.5 Observed waveform
3.8 DC Gain Adjustment on the Analog Assembly
NOTE Before performing this DC gain adjustment, the Flatness adjustment of the Attenuator
Board must have been completed.
Procedure
1) Remove the main unit cover and the two fixing plates..
2) Allow the unit to warm up for 15 minutes or more.
3) Connect each instrument as shown in Figure 3.6 Connection method.
CH 1 to 8
Conversion Adapter
DL2700
DC Voltage Standard
Cable
Figure 3.6 Connection method
4) Press the [INITIALIZE] key and select the “Initialize” option to execute initialization.
5) Press the [CAL] and “Calibration” option to perform calibration.
3.8 DC Gain Adjustment on the Analog Assembly